Fixing Bad ECG Tracings: Common Artifacts and Technical Solutions

High-quality ECGs require meticulous technique. Here's a systematic troubleshooting guide for common degrading factors.

Patient-Related Issues

  • Motion/tremor: Causes baseline wander, muscle artifact. → Relax patient, warm room, support limbs.
  • Poor skin prep: Oil, sweat, lotion, dead skin increase impedance. → Clean with alcohol, lightly abrade.
  • Body habitus: Obesity, large breasts, thick chest hair impair contact. → Shave, use conductive gel, consider alternative lead placement.

Electrode & Equipment Problems

  • Misplacement: Alters axis, mimics pathology. → Follow standard AHA positions.
  • Loose/expired electrodes: Dry gel, poor adhesion. → Use fresh electrodes, check expiration.
  • Lead reversal: LA/RA swap inverts lead I, swaps II/III. → Color-code check.
  • Cable/connector fault: Intermittent signal. → Inspect, replace.
  • Electrical interference: 50/60 Hz noise from outlets/equipment. → Ground patient, use battery mode, notch filter.

Settings Errors

  • Gain: 10 mm/mV standard; wrong gain distorts amplitude.
  • Speed: 25 mm/s standard.
  • Filter: 0.05–150 Hz diagnostic; 0.5–40 Hz monitor.

Systematic troubleshooting yields diagnostic-quality tracings.