Piercings and X-Ray Imaging: Safety, Artifacts, and Communication Protocols

Body piercings are common; their interaction with radiographic imaging is manageable with proper communication and technique.

Physics of Piercing Artifacts

  • Metallic jewelry (steel, titanium, gold, niobium): High atomic number → intense photoelectric absorption → radiopaque streaks/artifacts obscuring underlying anatomy.
  • Non-metallic (glass, acrylic, PTFE, bioplast): Radiolucent — minimal artifact.
  • Size/Geometry: Larger gauge, dangling jewelry = larger artifact volume.

Anatomic Considerations by Piercing Site

Piercing SiteImaging AffectedArtifact SeverityTypical Action
Ear (lobe, helix, tragus, conch)Skull, facial bones, cervical spine, dentalHigh (skull/dental)Remove for head/neck imaging
Nose (nostril, septum, bridge)Facial bones, sinuses, skull baseHighRemove for facial/sinus imaging
Lip/Labret/Monroe/MedusaMandible, dental, cervicalHigh (dental)Remove for dental/jaw imaging
TongueCervical spine, soft tissue neckModerateRemove for neck imaging
NippleChest, sternum, ribsModerateRemove for chest imaging
NavelAbdomen, pelvis, lumbar spineLow–ModerateRemove for abdominal/pelvic imaging
GenitalPelvis, hips, femursLow–ModerateRemove for pelvic imaging
Dermal/Microdermal (any site)Local areaVariable (anchor visible)Remove if in field of view

Clinical Decision Framework

  1. Patient informs tech: "I have piercings at [sites]."
  2. Tech assesses: Is jewelry in/adjacent to FOV (Field of View)? Is it ferromagnetic (MRI safety)?
  3. If in FOV: Request removal. Provide clean container, gloves, mirror.
  4. If removal impossible (fresh <8–12 wks, embedded, patient refusal):
    • Reposition patient to move jewelry out of FOV.
    • Use angled projections.
    • Lead shielding over jewelry only (reduces scatter, not primary beam artifact).
    • Note limitation on report: "Piercing artifact may obscure [anatomy]."
  5. New piercings: If removal risks closure/infection, discuss with radiologist. May accept limited study or alternative imaging (US, MRI if non-ferromagnetic).

MRI-Specific Concerns

  • Ferromagnetic jewelry: Projectile risk, torque, heating — MUST REMOVE.
  • Non-ferromagnetic (Ti, niobium, glass): Safe at 1.5T/3T but may cause susceptibility artifact.
  • Screening: MRI safety questionnaire + verbal confirmation + wand/ferromagnetic detector.

Radiation Protection

  • Jewelry does not increase patient dose significantly (scatter minimal).
  • Lead apron/shield for gonads/thyroid per standard protocol — independent of jewelry.

Best Practices

  • Pre-appointment reminder: "Remove all jewelry in area to be imaged."
  • Facility provides piercing removal kits (sterile gloves, hemostats, containers).
  • Document: "Piercings removed: [sites]" or "Piercings retained: [sites] — artifact noted."
  • Patient education: "Your piercing can hide a fracture. Removing it ensures we see everything."

Piercings are manageable with communication. The radiologist's goal: diagnostic images. The patient's goal: keep piercing intact. Usually both achievable.