Measuring Tidal Volume with a Spirometer: Step-by-Step Procedure

Tidal volume (VT) is the volume of air moved in/out during quiet breathing. Spirometry measures it via slow vital capacity (SVC) maneuver.

Equipment & Preparation

  • Calibrated spirometer (volume or flow type) with bacterial filter.
  • Clean mouthpiece; nose clips.
  • Subject seated, comfortable, dentures in if normally worn.

Procedure (SVC Method)

  1. Explain; demonstrate.
  2. Subject breathes tidally through mouthpiece to acclimate (3–4 breaths).
  3. At end of normal expiration (functional residual capacity), inspire slowly and maximally to total lung capacity (TLC).
  4. Without pause, expire slowly and completely to residual volume (RV).
  5. Repeat ≥3 acceptable maneuvers (AER/ERS criteria: volume reproducibility ≤0.15 L).

Deriving Tidal Volume

  • VT is not directly measured by SVC; it's obtained from quiet breathing trace (tidal breathing loop) recorded before/after SVC.
  • Alternatively, VT = Minute Ventilation (VE) / Respiratory Rate (fR) during quiet breathing.

Quality Control

  • Calibrate daily (3-L syringe).
  • Check for leaks, obstruction.
  • Coach for slow, steady flow; avoid glottis closure, cough.

Reference Values

VT ~6–8 mL/kg ideal body weight (≈400–600 mL adult). Varies with age, sex, height, posture, activity.