Diaphragmatic Attenuation Artifact in Cardiac Stress Imaging

Diaphragmatic attenuation artifact appears as a fixed inferior wall perfusion defect on SPECT/PET myocardial perfusion imaging (MPI) due to photon absorption by the diaphragm/liver.

Mechanism

Diaphragm moves with respiration. During acquisition, its position varies, causing inconsistent attenuation of ⁹⁹ᵐTc or ⁸²Rb photons from the inferior/apical myocardium. Mimics inferior ischemia.

Clinical Significance

  • Common (up to 30% of studies); usually benign.
  • Can obscure true inferior infarction or ischemia.
  • More prominent in men (larger liver), supine imaging, high BMI.

Mitigation Strategies

  • Respiratory gating: Acquire at end-expiration (diaphragm highest).
  • Attenuation correction (AC): CT-based (SPECT/CT, PET/CT) maps tissue density → corrects counts.
  • Prone imaging: Moves heart away from diaphragm.
  • Upright/sitting acquisition: Gravity pulls diaphragm down.
  • Alternative modalities: Cardiac MRI (no ionizing radiation), stress echo.

Interpreters must recognize artifact pattern (fixed, inferior, no matching wall motion abnormality on gated SPECT) and correlate with clinical probability.