Nasopharyngeal Airway (NPA) Sizing: Tip-to-Earlobe Measurement Technique

The standard method to select the correct length of a nasopharyngeal airway (NPA) is the tip-to-earlobe technique.

Procedure

  1. Place the flanged end of the NPA at the patient's naris (nostril opening).
  2. Extend the airway along the side of the face.
  3. Measure to the earlobe (caudal border of the pinna) on the same side.
  4. The distance from flange to earlobe approximates the distance from naris to nasopharynx.

Alternative / Confirmatory Methods

  • Tip-to-angle-of-mandible: Some protocols use mandibular angle as landmark.
  • Age/height-based sizing: Pediatric formulas (e.g., length = (age/2) + 12 cm) — less accurate.
  • Direct visualization: During insertion, advance until flange at naris; confirm position with laryngoscopy or fiberoptic scope.

Sizing Considerations

  • Too short: Tip in oropharynx → gagging, ineffective stenting.
  • Too long: Tip in esophagus or trachea → gastric distension, aspiration, laryngospasm.
  • Diameter: Largest that passes easily; typically 6–8 mm adult, 3–5 mm pediatric. French size = 3 × diameter (mm).

Key Points

  • Lubricate with water-soluble gel (not petroleum).
  • Insert along floor of nasal cavity (bevel toward septum).
  • Contraindicated in suspected basilar skull fracture (risk of intracranial placement).
  • Reassess position after any neck movement or patient transport.

Respiratory Disorders - Related Articles