Blood Pressure Measurement: Sphygmomanometer Technique Step-by-Step
Accurate BP measurement requires standardized technique (AHA/ESH guidelines).
Equipment
- Validated upper-arm cuff device (automated oscillometric preferred for routine).
- Manual: Aneroid sphygmomanometer + stethoscope (auscultatory – gold standard).
Patient Preparation
- Rest 5 min, seated, back supported, feet flat, legs uncrossed.
- No caffeine, exercise, smoking 30 min prior.
- Arm bare, supported at heart level.
Cuff Placement
- Correct size: Bladder encircles 80% arm, width 40% circumference.
- Lower edge 2–3 cm above antecubital fossa.
Manual Auscultatory Method
- Inflate to 20–30 mmHg above radial pulse disappearance.
- Deflate 2–3 mmHg/sec.
- Korotkoff Phase I (first tap) = Systolic.
- Phase V (disappearance) = Diastolic.
- Record to nearest 2 mmHg.
Automated Devices
Follow manufacturer; validate annually. Take 2–3 readings 1 min apart; average.
Common Errors
- Wrong cuff size (too small → falsely high).
- Arm below heart level.
- Talking/moving during measurement.
- Terminal digit preference (rounding to 0/5).