Correct Dipstick Urinalysis Technique: Dipping to the Indicated Line for Accuracy
Why the Fill Line Matters
- Reagent pads are calibrated for a specific urine volume.
- Under-dipping: Insufficient wetting → false-negative (color development incomplete).
- Over-dipping: Excess urine → reagent leaching, cross-contamination between pads, false-high readings.
Standard Procedure
- Mix urine gently; test within 30 min (or refrigerate).
- Dip strip completely to the marker line (usually 1–2 cm from tip).
- Withdraw immediately; drag edge against container rim to remove excess.
- Hold horizontally; read at exact times per manufacturer (e.g., 30 s, 60 s, 120 s).
- Compare to color chart under good lighting.
Quality Control
- Run positive/negative controls each new bottle & daily.
- Store strips capped, desiccated, 15–30°C.
- Check expiration date.